Q. Why am I getting as high as 170% recovery when using Total Organic Carbon (TOC) with Alconox detergent?

A. There are several possible reasons for high TOC recovery:

1. TOC contamination from sample vials – be sure to use TOC grade to clean sample vials – be sure to use equipment blanks to compensate for any interference from the sample vials.

2. Be sure to use TOC grade swabs for your TOC testing.

Do you have a question for our experts about critical cleaning? Please visit Ask Alconox. Or to test an Alconox, Inc. detergent for free, please complete the questionnaire at Get Sample. For additional technical information on cleaning validation, please Download our Pharmaceuitcal and Medical Device Cleaning Validation References PDF.

 